TH0530


• Available in 1 kA and 2 kA current specifications

• Measurement frequency: DC ~ 1 kHz

• Accuracy class: 0.005 / class 0.01

• Through-core and direct-in measurement

• The aperture of Ф 70 mm facilitates penetration of high-current wires

• Analog I/V output mode is supported

TH0530  is a wide measurement range, high measurement accuracy, small and portable ammeter, using broadband current comparator technology, suitable for measuring broadband high current in the field, or for calibrating AC and DC high current sources and ammeters.

Power supply

AC:85 V ~ 265 V / 47 Hz ~ 63 Hz

DC:120 V ~ 370 V

Maximum power consumption

100 VA

Warm-up time

30 minutes

Working environment

Temperature: 0°C ~ 50°C

Humidity: 30% R·H ~ 80%R·H, no condensation.

Others: No electromagnetic field interference.

Storage environment

Temperature: -30°C ~ 70°C

Humidity: 10% R· H ~ 90%R·H, no condensation.

Altitude

0 m ~ 3000 m

Communication interface

RS 232 × 1

Size

175 mm (W) × 330 mm (D)  × 275 mm (H)

Display screen

5.6-inch color LCD display

Quality

Approx 13 kg

 

 AC direct current measurement

Measurement

manner

Current range

Resolution

Frequency

Hz

Measurement uncertainty (k=2)

(23°C ± 5°C)

Temperature coefficient (/°C).

(0°C~50°C)

TH0530

-1 kA

TH0530

-2 kA

Class 0.005

Class

0.01

Class 0.005

Class

0.01

(ppm*RD+ppm*RG)[1]

Straight-in

2.5 A

5 A

1 μA

DC

40 + 10

80 + 20

3 + 2

6 + 4

40≤F≤400

150 + 50

220 + 80

400<F≤1k

350 + 150

350 + 150

10 A

20 A

10 μA

DC

40 + 10

80 + 20

3 + 2

6 + 4

40≤F≤400

150 + 50

220 + 80

400<F≤1k

350 + 150

350 + 150

Piercing

250 A

500 A

100 μA

DC

40 + 10

80 + 20

3 + 2

6 + 4

40≤F≤400

150 + 50

220 + 80

400<F≤1k

350 + 150

350 + 150

1 kA

2 kA

1 mA

DC

40 + 10

80 + 20

3 + 2

6 + 4

40≤F≤400

150 + 50

220 + 80

400<F≤1k

350 + 150

350 + 150

Note [1]: RD is the reading value and RG is the range value.

• Measurement up to 110%* measuring range

• Measurement frequency: DC ~ 1 kHz

• Range switching: manual/automatic gear shifting

• Display digits: 7 digits, decimal

 



 

• Formation of AC and DC current measurement system

• DC stable power supply test

• Battery charge and discharge test and calibration

• Calibrate the multifunction calibrator

• Calibrate the AC and DC current source/meter

• Clamp meter calibration device traceability

• Voltage output reference

• Ripple and distortion measurement