TD2600


• DCV measurement range: 100 V ~ 5 kV.

• Accuracy class: 0.02, 0.05.

• Max. range 10 kV / 20 kV / 50 kV are optional.

• Stability testing for HV source, data analysis.

• Short circuit and overload protection.

• LCD touch screen

• RS232 interface.

TD2600 is a wide range, high precision DC high voltage meter. It’s applied for calibrating DC high voltage source or meter, high voltage probe, etc.

5. Specifications

Model

Range

(kV)

Accuracy

±(ppm of reading + ppm of range) [1]

Temperature Coefficient (±ppm/K)

Class 0.05

Class 0.02

Class 0.05

Class 0.02

TD2600-x-5kV

1、2、5

300 + 200

120 + 80

< 15

< 10

TD2600-x-10kV

1、2、5、10

300 + 200

120 + 80

< 15

< 10

TD2600-x-20kV

2、5、10、20

300 + 200

120 + 80

< 15

< 10

TD2600-x-50kV

51020、50

300 + 200

-

< 15

-

Note [1] : (ppm = parts per million) (e.g.,10ppm = 0.001%).

• Measurement range :(10% ~ 100%)*range, 6 digits display,

• Manual or automatic range switch


6. General Specifications

Power Supply

AC (220 ± 22) V,(50 ± 2) Hz

Temperature

Working temperature:0℃~45℃;

Storage temperature:-20℃~70℃

Humidity

Working humidity:< 80% @ 30℃,< 70% @ 40℃,< 40% @ 50℃

Storage humidity:(20%~80%) R·H,non-condensing

Interface

RS232 interface

 Calibration for DC High Voltage Source / Meter

• DCV measurement: 100 V ~ 5 kV / 10 kV / 20 kV / 50 kV

• Apply for calibrating DC high voltage source or meter, high voltage probe, etc