TD1540
• Accuracy: class 0.02 / 0.05
• DC standard current:0.5 A~600 A
• Precision DC voltage measurement: 100 μV~11 V
• Resistance measurement range: 200 nΩ~6 Ω
• A variety of quantity output adjustment methods
• Measuring the resistance value and basic error of the shunt, drawing R (I) and R (t) curve
• Professional test software (option)
Power supply | AC ( 220±22 ) V,( 50±2 ) Hz; |
Warm up time | 2h; |
Maximum power consumption | 4.8 kVA |
Sampling rate | 0.5 sps |
Temperature performance | Operating temperature: 15 °C~30 °C; Storage temperature: -20 °C ~ 70 °C; |
Humidity performance | Operating humidity: ( 20 % ~ 50% ) R·H, No condensation; Storage humidity: ( 15 % ~ 80% ) R·H, No condensation; |
Altitude | < 3000 m |
Weight | About 49.5 kg |
Communication interface | RS 232 |
Unit size | 623 mm (W)×548 mm (D) × 260 mm (H) (Excluding handle and support foot) |
5.1 DC current output DCI
• Output range:0.5A~600 A, Ripple coefficient: < 0.5 %
• 7-digits display, Regulating fineness: 0.001%*RG
• Protection function: Open circuit protection, Overload protection
5.2 DC small signal voltage measurement DCV
• Measuring range: ±(100 μV~11 V), Manual/automatic range switching, 7-digits display
• Input resistance: >1 GΩ, Input protection: 50 Vpk, continuously