TD1855 Multifunction Calibrator


• DC voltage output: 20 mV ~ 1100 V

• DC current output: 2 μA ~ 22 A33 A is optional

• AC voltage output: 20 mV ~ 1100 V

• AC current output: 200 μA ~ 22 A33 A is optional

• Sinusoidal wave frequency: 45 Hz ~ 1100 Hz

• Simulated DC resistance: 10 Ω ~ 220 MΩ

• 2 MHz square wave frequency output

• Phase output

• AC/DC power output

• Remote adjustment (optional)

• Clamp meter calibration (optional)

• Transducer calibration(optional)

•  5.6-inch Color Touch Screen

TD1855 is a multi-function calibrator, integrates AC\DC voltage source, AC\DC current source, DC resistance source, etc. The accuracy class is 0.05. It’s applied to calibrating 3.5 digits multimeter, voltmeter, ammeter, ohmmeter, power meter.

1 DC Voltage/Current Output

Range

Resolution

Stability (±%/min)

Accuracy

±(ppm*RD+ppm*RG)[1]

Max burden (mA)

Ripple Factor(%)

200 mV

1 μV

0.01

300 + 200

300

<0.5

2 V

10 μV

0.01

300 + 200

500

<0.5

10 V

100 μV

0.01

300 + 200

500

<0.5

30 V

100 μV

0.01

300 + 200

800

<0.5

100 V

1 mV

0.01

300 + 200

300

<0.5

300 V

1 mV

0.01

300 + 200

100

<0.5

600 V

1 mV

0.01

300 + 200

50

<0.5

1000 V

10 mV

0.01

300 + 200

30

<0.5

 

Range

Resolution

Stability

(±%/min)

Accuracy

±(ppm*RD+ppm*RG)

Compliance Voltage (V)

Ripple Factor(%)

20μA

100pA

0.01

300 + 200

10

<0.5

200μA

1 nA

0.01

300 + 200

9

<0.5

2mA

10 nA

0.01

300 + 200

9

<0.5

20mA

100 nA

0.01

300 + 200

9

<0.5

200 mA

1 μA

0.01

300 + 200

9

<0.5

2 A

10 μA

0.01

300 + 200

3.5

<0.5

10 A

100 μA

0.01

300 + 200

2.5

<0.5

20 A (30 A)[2]

100 μA

0.01

300 + 200

2.5

<0.5

Note[1]: RD is the reading value, RG is the range value;

Note[2]: The 30 A range is optional.

• Voltage output range: 20 mV ~ 1100 V

• Current output range: 2 μA ~ 22 A\33 A

• Protections: voltage source short-circuit protection, current source open-circuit protection, overload protection

 

2 AC Voltage/Current Output

Range

Resolution

Stability

(±%/min)

Accuracy

±(ppm*RD+ppm*RG)

Max Burden (mA)

Harmonic (%)

200 mV

1 μV

50 μV

300 + 60 μV

500

<1.0

2 V

10 μV

0.01

300 + 200

500

<0.5

10 V

100 μV

0.01

300 + 200

800

< 0.5

30 V

100 μV

0.01

300 + 200

800

< 0.5

100 V

1 mV

0.01

300 + 200

300

< 0.5

300 V

1 mV

0.01

300 + 200

100

< 0.5

600 V

1 mV

0.01

300 + 200

50

< 0.5

1000 V

10 mV

0.01

300 + 200

30

< 0.5

 

Range

Resolution

Stability

(±%/min)

Accuracy

±(ppm*RD+ppm*RG)

Compliance Voltage (V)

Harmonic (%)

2 mA

10 nA

0.6 μA

300 +0.6 μA

50

<0.5

20 mA

100 nA

0.01

300 + 200

50

<0.5

200 mA

1 μA

0.01

300 + 200

50

<0.5

1 A

10 μA

0.01

300 + 200

15

<0.5

2 A

10 μA

0.01

300 + 200

10

<0.5

5 A

10 μA

0.01

300 + 200

5

<0.5

20 A(30 A)[1]

100 μA

0.01

300 + 200

2.5

<0.5

Note [1]: 30 A range is optional.

• Voltage output range: 20 mV ~ 1100 V

• Current output range: 2 μA ~ 22 A\33 A

• Voltage and current output frequency: 45 Hz ~ 1100 Hz

• Protections: voltage source short-circuit protection, current source open-circuit protection, overload protection

 

3 Simulated DC Resistance

Range

Resolution

Accuracy

± (ppm*RD+ppm*RG)

Allowable Current

100 Ω

1 mΩ

300 + 200

1mA80 mA

300 Ω

1 mΩ

300 + 200

1mA80 mA

1 kΩ

10 mΩ

300 + 200

100μA8 mA

3kΩ

10 mΩ

300 + 200

100μA8 mA

10 kΩ

100 mΩ

300 + 200

10μA0.8 mA

30 kΩ

100 mΩ

300 + 200

10μA0.8 mA

100 kΩ

1 Ω

300 + 200

10μA0.8 mA

300 kΩ

1 Ω

300 + 200

1μA80 μA

1 MΩ

10 Ω

300 + 200

1μA80 μA

10 MΩ

100 Ω

600 + 400

100nA8 μA

100 MΩ

1000 Ω

3000 + 2000

10nA0.8 μA

200 MΩ

1000 Ω

3500 + 2500

10nA0.8 μA

• Output range: 10 Ω ~ 220 MΩ

• Over-current and reverse connection protection

 

4 Sinusoidal Wave Frequency Output

Output Range[1]

Resolution

Accuracy

45.0000 Hz ≤ F ≤ 99.9999 Hz

0.0001 Hz

±0.01%

100.000 Hz ≤ F ≤ 999.999 Hz

0.001 Hz

±0.01%

1000.00 Hz ≤ F ≤ 1100.00 Hz

0.01 Hz

±0.01%

Note [1]: The output mode is AC voltage or AC current.

 

5 AC/DC Power Output

Frequency Range (f)

Voltage Range (U)

Current Range (I)

Optimum Accuracy

%*power output [1][2]

DC

20 mV ~ 1100 V

2 μA ~ 22 A/33 A

±0.1

45 Hz ~ 65 Hz

3 V ~ 600 V

0.2 mA ~ 22 A/33 A

±0.1

 

6 Phase and Power Factor Output45 Hz65 Hz

Phase

Power Factor

Accuracy

WATTS

VARS

λ=cosφ

Phase

Power Factor

90°

1.0000

0.1°

0.000%

10°

80°

0.9848

0.1°

0.031%

20°

70°

0.9397

0.1°

0.064%

30°

60°

0.8660

0.1°

0.101%

40°

50°

0.7660

0.1°

0.147%

50°

40°

0.6429

0.1°

0.208%

60°

30°

0.5000

0.1°

0.302%

70°

30°

0.3420

0.1°

0.480%

80°

10°

0.1737

0.1°

0.990%

90°

0.0000

0.1°

Note [1]: Phase range: 0.000 °~ 359.999 °, fineness: 0.005 °

Note [2]: Power factor range: -1 ~ 0 ~ 1, fineness: 0.000 1

Note [3]: Calculation formulas for the accuracy of other power factor:

 

7 Pulse Frequency Output

Range[1]

Resolution

Accuracy

Shake

1.000 00 Hz ≤ F ≤ 9.999 99 Hz

10 μHz

± (20 ppm*RD + 20 μHz)

<20 ns

10.000 0 Hz ≤ F ≤ 99.999 9 Hz

0.1 mHz

100.000 Hz ≤ F ≤ 999.999 Hz

1 mHz

1.000 00 kHz ≤ F ≤ 9.999 99 kHz

10 mHz

10.000 0 kHz ≤ F ≤ 99.999 9 kHz

0.1 Hz

100.000 kHz ≤ F ≤ 999.999 kHz

1 Hz

1.000 00 MHz ≤ F ≤ 2.000 00 MHz

10 Hz

Note [1]: The output type is TTL level.

 

8 DC Small-signal Measurement (optional)

Type

Range

Range

Accuracy

DCV

1 V

± (01.2) V

±(0.006%*RD + 0.004%*RG)

10 V

± (012) V

±(0.006%*RD + 0.004%*RG)

DCI

2 mA

± (02.4) mA

±(0.006%*RD + 0.004%*RG)

20 mA

± (024) mA

±(0.006%*RD + 0.004%*RG)