TD1850 Multifunction Calibrator
• DC
voltage output: 20 mV ~ 1100 V
• DC current
output: 2 μA ~ 22 A(33
A is optional)
• AC voltage
output: 20 mV ~ 1100 V
• AC
current output: 200 μA ~ 22 A(33
A is optional)
• Sinusoidal wave
frequency: 45 Hz ~ 1100 Hz
• Simulated DC
resistance: 10 Ω ~ 220 MΩ.
• 2 MHz square
wave frequency output
• Remote adjustment
(optional)
• Clamp meter
calibration (optional)
• Transducer calibration(optional)
• 7.0-inch
Color Touch Screen
TD1850 is a multi-function calibrator, integrates AC\DC
voltage source, AC\DC current source, DC resistance source,
etc. The accuracy class is 0.05. It’s applied to
calibrating 3.5 digits multimeter, voltmeter, ammeter, ohmmeter, power
meter.
1 DC Voltage/Current Output
Range |
Resolution |
Stability (±%/min) |
Accuracy ±(ppm*RD+ppm*RG)[1] |
Max
Burden (mA) |
Ripple
Factor(%) |
200 mV |
1 μV |
0.01 |
300 + 200 |
300 |
<0.5 |
2 V |
10 μV |
0.01 |
300 + 200 |
500 |
<0.5 |
10 V |
100 μV |
0.01 |
300 + 200 |
500 |
<0.5 |
30 V |
100 μV |
0.01 |
300 + 200 |
800 |
<0.5 |
100 V |
1 mV |
0.01 |
300 + 200 |
300 |
<0.5 |
300 V |
1 mV |
0.01 |
300 + 200 |
100 |
<0.5 |
600 V |
1 mV |
0.01 |
300 + 200 |
50 |
<0.5 |
1000 V |
10 mV |
0.01 |
300 + 200 |
30 |
<0.5 |
Ranges |
Resolution |
Stability (±
%/min) |
Accuracy ±
(ppm*RD+ppm*RG) |
Compliance
Voltage (V) |
Ripple
Factor(%) |
20 μA |
100pA |
0.01 |
300 + 200 |
10 |
<0.5 |
200 μA |
1 nA |
0.01 |
300 + 200 |
9 |
<0.5 |
2 mA |
10 nA |
0.01 |
300 + 200 |
9 |
<0.5 |
20 mA |
100 nA |
0.01 |
300 + 200 |
9 |
<0.5 |
200 mA |
1 μA |
0.01 |
300 + 200 |
9 |
<0.5 |
2 A |
10 μA |
0.01 |
300 + 200 |
3.5 |
<0.5 |
10 A |
100 μA |
0.01 |
300 + 200 |
2.5 |
<0.5 |
20 A
(30 A)[2] |
100 μA |
0.01 |
300 + 200 |
2.5 |
<0.5 |
Note[1]: RD
is the reading value, RG is the range value; Note[2]: The
30 A range is optional. |
• Voltage output range: 20 mV ~ 1100 V
• Current output range: 2 μA ~ 22 A\33 A
• Protections: voltage source short-circuit
protection, current source open-circuit protection, overload
protection
2 AC Voltage/Current Output
Range |
Resolution |
Stability ( ±
%/min) |
Accuracy ±
(ppm*RD+ppm*RG) |
Max
Burden (mA) |
Harmonic
(%) |
200 mV |
1 μV |
50 μV |
300 + 60
μV |
500 |
<1.0 |
2 V |
10 μV |
0.01 |
300 + 200 |
500 |
<0.5 |
10 V |
100 μV |
0.01 |
300 + 200 |
800 |
< 0.5 |
30 V |
100 μV |
0.01 |
300 + 200 |
800 |
< 0.5 |
100 V |
1 mV |
0.01 |
300 + 200 |
300 |
< 0.5 |
300 V |
1 mV |
0.01 |
300 + 200 |
100 |
< 0.5 |
600 V |
1 mV |
0.01 |
300 + 200 |
50 |
< 0.5 |
1000 V |
10 mV |
0.01 |
300 + 200 |
30 |
< 0.5 |
Range |
Resolution |
Stability ( ±
%/min) |
Accuracy ±
(ppm*RD+ppm*RG) |
Compliance
Voltage (V) |
Harmonic
(%) |
2 mA |
10 nA |
0.6 μA |
300 +0.6 μA |
50 |
< 0.5 |
20 mA |
100 nA |
0.01 |
300 + 200 |
50 |
< 0.5 |
200 mA |
1 μA |
0.01 |
300 + 200 |
50 |
< 0.5 |
1 A |
10 μA |
0.01 |
300 + 200 |
15 |
< 0.5 |
2 A |
10 μA |
0.01 |
300 + 200 |
10 |
< 0.5 |
5 A |
10 μA |
0.01 |
300 + 200 |
5 |
< 0.5 |
20 A(30 A)[1] |
100 μA |
0.01 |
300 + 200 |
2.5 |
< 0.5 |
Note [1]: 30 A
range is optional. |
• Voltage output range: 20 mV ~ 1100 V
• Current output range: 2 μA ~ 22 A\33 A
• Voltage and current output frequency: 45
Hz ~ 1100 Hz
• Protections:
voltage source short-circuit protection,
current source open-circuit protection, overload protection
3 Simulated DC Resistance
Range |
Resolution |
Accuracy ±
(ppm*RD+ppm*RG) |
Allowable
Current |
100 Ω |
1 mΩ |
300 + 200 |
1mA
~ 80 mA |
300 Ω |
1 mΩ |
300 + 200 |
1mA
~ 80 mA |
1 kΩ |
10 mΩ |
300 + 200 |
100μA
~ 8 mA |
3kΩ |
10 mΩ |
300 + 200 |
100μA
~ 8 mA |
10 kΩ |
100 mΩ |
300 + 200 |
10μA
~ 0.8 mA |
30 kΩ |
100 mΩ |
300 + 200 |
10μA
~ 0.8 mA |
100 kΩ |
1 Ω |
300 + 200 |
10μA ~ 0.8
mA |
300 kΩ |
1 Ω |
300 + 200 |
1μA
~ 80 μA |
1 MΩ |
10 Ω |
300 + 200 |
1μA
~ 80 μA |
10 MΩ |
100 Ω |
600 + 400 |
100nA
~ 8 μA |
100 MΩ |
1000 Ω |
3000 + 2000 |
10nA
~ 0.8 μA |
200 MΩ |
1000 Ω |
3500 + 2500 |
10nA
~ 0.8 μA |
• Output range: 10 Ω ~ 220 MΩ
• Over-current and reverse connection protection
4 Sinusoidal Wave Frequency
Output
Output Range[1] |
Resolution |
Accuracy |
45.0000 Hz ≤ F
≤ 99.9999 Hz |
0.0001 Hz |
±0.01% |
100.000 Hz ≤ F
≤ 999.999 Hz |
0.001 Hz |
±0.01% |
1000.00 Hz ≤ F
≤ 1100.00 Hz |
0.01 Hz |
±0.01% |
Note [1]: The
output mode is AC voltage or AC current. |
5 Pulse Frequency Output
Range[1] |
Resolution |
Accuracy |
Shake |
1.000 00 Hz ≤ F
≤ 9.999 99 Hz |
10 μHz |
±(20 ppm*RD +
20 μHz) |
<20 ns |
10.000 0 Hz ≤ F
≤ 99.999 9 Hz |
0.1 mHz |
||
100.000 Hz ≤ F
≤ 999.999 Hz |
1 mHz |
||
1.000 00 kHz ≤
F ≤ 9.999 99 kHz |
10 mHz |
||
10.000 0 kHz ≤
F ≤ 99.999 9 kHz |
0.1 Hz |
||
100.000 kHz ≤ F
≤ 999.999 kHz |
1 Hz |
||
1.000 00 MHz ≤
F ≤ 2.000 00 MHz |
10 Hz |
||
Note [1]: The
output type is TTL level. |
6 DC Small-signal Measurement (optional)
Type |
Range |
Measurement
range |
Accuracy |
DCV |
1 V |
±(0 ~ 1.2) V |
±(0.006%*RD +
0.004%*RG) |
10 V |
±(0 ~ 12) V |
±(0.006%*RD +
0.004%*RG) |
|
DCI |
2 mA |
±(0 ~ 2.4) mA |
±(0.006%*RD +
0.004%*RG) |
20 mA |
±(0 ~ 24) mA |
±(0.006%*RD +
0.004%*RG) |