TH0110


• Output voltage range: ± (0 ~ 10 V).

• Reference voltage output: 2.5 V / 4.096 V / 5 V / 10 V

• Typical short-term stability: 0 2 ppm/min

• Resolution: 1 nV (8 bits).


TH0110 is an ultra-precise DC voltage reference with high stability, high accuracy, low temperature drift and small size.

Power supply

Power supply: AC ( 220 ± 22 ) V, ( 50 ± 2 ) Hz;

Warm-up time

Not less than 1 hour, and the warm-up time again after shutting down in the middle should not be less than 2 times the shutdown time;

Temperature performance

Working temperature: 15 °C ~ 35 °C;

Storage temperature: -40 °C ~ 50 °C;

Humidity performance

Operating humidity:( 0 ~ 90% ) R·H @ ≤ 28°C, no condensing;

≤80%R·H @ ≤ ( 28°C ~ 35°C ),no-condensing;

Storage humidity: ( 0 ~ 90% ) R·H, no condensing.

Altitude

< 3000 m

Range

Resolution

Annual uncertainty (k=2).

± (ppm* setpoint + μV).

10 mV

1 nV

15 + 0. 12

100mV

10 nV

7.5 + 0.4

1 V

100 nV

5.0 + 0. 7

10 V

1 μV

3.0 + 2.5

• Output range: ± (0 ~ 10 V).

• Manual/automatic shifting, 8-digit display

• Detect ADC chip linearity

• Calibrate a high-precision DC voltmeter

• Calibrate the digital nanovoltmeter

• Calibrate the potentiometer