Single Phase Portable Meter Test System MCCS1.1
Small size and weight.
Used in lab or worksite.
LCD display.
Accuracy class: 0.1 (direct mode); 0.2 (clamp mode).
Measuring range: upto 300V /100A (direct mode).
Power and Energy test, Demand test, Error test, Dial test, Start, Creep test.
Support RS232 communication with PC to upload the test data.
Accuracy class of direct mode | 0.1 |
Accuracy class (clamp 10A, 50A, 100A, 300A…optional) | 0.2 |
Technical index of in-built single phase standard meter | |
Class | 0.1 |
Input current range | 0.025A, 0.05A, 0.1A, 0.25A, 0.5A, 1A, 2.5A, 5A, 10A, 20A, 50A, 100A |
Input voltage range | up to 300V |
Display resolution | 6 digits (valid) |
Output voltage | up to 240V, 120% adjustable (maximum up to 288V) |
Adjust resolution | better than 0.01% |
Output current range | 0.05A, 0.25A, 1A, 5A, 20A, 100A |
0-120% adjustable; | |
Adjust resolution | better than 0.01% |
Minimum output current | 1mA (≤5%) |
Output power | 30VA for current and 50VA for voltage (only test one meter) |
100VA for current and 50VA for voltage (test three meters at the same time) | |
Output stability of voltage, current and power | ≤0.05% /120s |
Waveform distortion of output voltage/current | ≤0.5% |
Frequency range | 45-65HZ |
Adjust resolution | 0.01Hz |
Phase angle range | -180.0o ~180.0o |
Adjust resolution | 0.01° |
Power supply | AC 220V±10%,50/60HZ; or from measuring circuit |
Power consumption | 200VA (maximum) |
Dimension & Weight | 361*236*163mm & 7kg (main unit) |