Additel 875-155 Dry Well Calibrator

If you are serious about portable temperature calibration tools, then you know a good dry well calibrator is more than just a stable heat source. The Additel 875 Series Dry Well Calibrators combine excellent performance in stability, radial and axial uniformity. And loading with speed, ruggedness and portability. But we don’t stop there! The Process Calibrator option adds the capabilities of a three-channel thermometer readout and a documenting process calibrator. We’ve also incorporated a unique option to select your own temperature range within the range of the model selected. We’re calling this the CYOR option or Choose Your Own Range option. When you purchase the CYOR option, you pick the upper and lower temperature range needed and we calibrate and optimize the dry well’s performance over your selected range. Each unit has a color touch screen display, dual-zone control, and much more. You are just going to love these new dry wells!

Process Calibrator Option

Each model offer has a Process Calibrator(PC)option. This process calibrator option combines the many features found in a HART documenting process calibrator with the temperature dry well. This option includes the ability to measure a reference PRT and two device under test channels which can measure, mA, voltage, switch, RTD or thermocouple. In addition to these measurement functions, this calibrator has full documenting capability of creating tasks, saving as found and as left results, and HART communication. The snap shot feature allows you to capture all information displayed on the screen with the push of a button. This unit also allow for data logging of all channels on an auto step function and a ramp function. By utilizing the reference PRT, you can select to control to the dry well set point using the internal sensor or the external reference PRT.

Self Calibration

We believe using an external reference probe as your standard is the best way to perform your temperature calibration. But we also recognize this method is not always necessary or convenient and depending on the application, using the internal control sensor would be preferred. Traditionally, the internal control sensor has a wide accuracy which can largely be contributed to its long-term drift. We’ve built-in a self calibration feature allowing you to run an automated calibration of the internal control sensor using your external reference. With literally a few selections the calibration will run automatically giving you a fresh, traceable calibration of the control sensor which will improve its accuracy as you will not have to account for its long term drift when used as the reference.

Short Probe Kit

The Additel 110 series short probe calibration kit is designed to help users of our ADT875-155 (-40°C dry well) calibrate temperature probes and transmitters with short probe lengths. This all-inclusive kit comes with everything needed to perform more accurate and reliable calibration for those challenging short probes. Choose between our metric or imperial kit to fit your needs. For more information, please watch our instructional short probe video found at www.additel.com

2.png

Specification875-155875-350875-660
Temperature Range at 23℃-40°C to 155°C33°C to 350°C33°C to 660°C
Display Accuracy±0.18°C at Full Range±0.2°C at Full Range±0.3°C at 33°C
±0.3°C at 420°C
±0.5°C at 660°C
Stability (30 min)±0.01°C at Full Range±0.02°C at Full Range±0.02°C at 33°C
±0.03°C at 50°C
±0.04°C at 420°C
±0.04°C at 660°C
Axial Uniformity at 60 mm (2.4 in)±0.07°C at Full Range±0.04°C at 33°C±0.05°C at 33°C
±0.1°C at 200°C±0.3°C at 420°C
±0.2°C at 350°C±0.5°C at 660°C
Radial Uniformity±0.01°C at Full Range±0.01°C at 33°C±0.02°C at 33°C
±0.015°C at 200°C±0.05°C at 420°C
±0.02°C at 350°C±0.1°C at 660°C
Loading Effect±0.1°C (Display Sensor)±0.15°C (Display Sensor)±0.15°C (Display Sensor)
±0.02°C (External Sensor)±0.015°C (External Sensor)±0.025°C (External Sensor)
Hysteresis (Display Sensor)0.025°C0.03°C0.1°C
Environmental Conditions8°C to 38°C guaranteed accuracy
0°C to 50°C, 0% to 90% RH non-condensing, 3000 M altitude for normal operation
Storage Conditions-20°C to 60°C
IP RatingIP20
Immersion Depth150 mm (5.9 in)
Insert OD25.8 mm (1.02 in)24.8 mm (0.98 in)
Heating Time13 min: -40°C to 155°C5 min: 33°C to 350°C15 min: 33°C to 660°C
5 min: -40°C to 23°C
8 min: 23°C to 155°C
Cooling Time28 min: 155°C to -40°C15 min: 350°C to 100°C23 min: 660°C to 100°C
8 min: 155°C to 23°C10 min: 100°C to 50°C12 min: 100°C to 50°C
20 min: 23°C to -40°C10 min: 50°C to 33°C12 min: 50°C to 33°C
Typical Time to Stability10 min
Resolution0.01°C
Units°C, °F, and K
Display6.5 in (165 mm) color touch screen
Size (H x W x D)320 x 170 x 330 mm (12.6 x 6.7 x 13.0 in)
Weight9.9 kg (21.8 lbs)8.2 kg (18.1 lbs)
Power Requirements90-254 VAC, 45-65 Hz, 580 W90-254 VAC, 45-65 Hz, 1200 W
Mechanical TestingVibration: 2 g (10-500 Hz), 30 min for 2 sides
Impact: 4 g three times
Drop test: 500 mm (19.6 in)
CommunicationUSB A, USB B, RJ45, WiFi, Bluetooth
LocalizationEnglish, Chinese, Japanese, Russian, German, French, Italian, and Spanish
Warranty1 year

Documents:

File nameUpload DateSize
Additel 875 Dry Well Calibrator Datasheet 2018.5.4 5 MB
Additel 875 Dry Well Calibrator User Manual 2019.4.2 4.2 MB
Additel Short Probe Temperature Calibration Kit Datasheet 2019.6.19 0.5 MB

Update Packages:

File name Upload Date Size
Additel 875 Firmware Update Package 2020.2.18 11 MB
HART Temperature Library Update Package 2019.6.5 50.7 MB
HART Pressure Library Update Package 2019.6.5 138.4 MB